Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
Input ABCD (AB) Position
Input ABCD (AB) Volt Gain
To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
The adjustment steps for channel C and D are only for the models
190-104 and 190-204.
PM9093
A
METER
2 CHANN. + METER
F2
F3
Input ABCD (AB) Volt Gain
F2
PM9092
PM9092
B
B
A
C
D
4 CHANNELS
Figure 5-16. Test Tool Input ABCD to 5502A Normal Output
F3
:READY
Warning
Note
PM9091
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
NO
50 OHM TERMINATION
NORMAL
SCOPE
PM9081
FLUKE 5502A CALIBRATOR
:READY
CL 0619.