Fluke ScopeMeter 190 II Series User Manual page 194

Hide thumbs Also See for ScopeMeter 190 II Series:
Table of Contents

Advertisement

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
Input A, B, C, D AC Voltage Accuracy Test (LF)
Bandwidth Test
To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
Reading 1, on A, V ac
Reading 2, on B, V ac
Reading 3, on C, V ac
Reading 4, on D, V ac
The test steps for channel C and D are only for the models
190-104, 190-204, and 190-504.
Input B AC Voltage Accuracy (HF) & Bandwidth Test
SAVE
F2
ENTER
A
Attenuator: Normal | Bandwidth: 10 kHz
B
Attenuator: Normal | Bandwidth: 10 kHz
C
Attenuator: Normal | Bandwidth: 10 kHz
D
Attenuator: Normal | Bandwidth: 10 kHz
CLEAR
SCOPE
F2
Note
Input A AC Voltage Accuracy (HF) &
Warning
Creating a Standard Test Tool Setup
SETUP
F4
| Bandwidth: 20 MHz
F4
| Bandwidth: 20 MHz
F4
| Bandwidth: 20 MHz
F4
| Bandwidth: 20 MHz
F1
ENTER
20 kHz
20 kHz
20 kHz
20 kHz