Technical data
8.2 CD 6 optics housing for use in hazardous areas
Climatic conditions
Ambient temperature
- during operation
- during transportation and storage
Atmospheric pressure
Humidity
8.2

CD 6 optics housing for use in hazardous areas

LDS 6, Optics Housing CD 6 for use in
hazardous areas
Dimension, optics housing
Dimension, purging tube
Weight
Power supply
Power consumption
Degree of protection
EMC - Electromagnetic compatibility
Electrical safety
Permissible ambient temperature (oper‐
ation) for the components of the Ex-rel‐
evant IP area
Permissible ambient temperature (stor‐
age and transport)
Atmospheric pressure
Maximum allowable work pressure
(MAWP)
Permissible ambient humidity
Optics housing/process interface
Measurement path length
Dust load
See also
Safe installation (Page 24)
60
 
5 ... 45 °C (41 ... 113 °F)
-40 ... +70 °C (-40 ... +158 °F)
800 ... 1 100 hPa (12 ... 16 psi)
<85 % relative humidity, non-condensing
Specification
L: 450 mm (17 3/4"), Ø: 163 mm (6 7/16"),
flange diameter DN80/PN16: 200 mm (7 7/8") or ANSI 4"/150
lbs.: 230 mm (9")
L: 400, 800 or 1 200 mm (15 3/4, 31 1/2 or 47 1/4"), Ø: 44 mm
(1 3/4")
2 x 11 kg (24 lbs.)
According to the intrinsically-safe standards (IEC/
EN 60079-11)
Max. 0.6 W
IP65 according to EN 60529
According to EN 61326 and standard classification of NA‐
MUR NE21
According to EN 61010-1, overvoltage classification II.
-20 ... +60 °C (-4 ... +140 °F)
-30 ... +70 °C (-22 ... +158 °F)
800 ... 1100 hPa (12 ... 16 psi)
For pressure-resistant window flange 1.6 MPa (232 psi) at
20 °C (68 °F)
For standard flange 1 050 hPa (15 psi) at 20 °C (68 °F)
<95 % relative humidity, non-condensing
DN65/PN6 and ANSI 4"/150 lbs.;
DN80/PN16 (only as pressure-resistant variant)
0.3 ... 12 m (1 ... 39 ft) depending on measuring conditions
(dust load)
Depending on particle size distribution and optical path length
LDS 6 & Optics housing CD 6 Compact Operating Instructions
Compact Operating Instructions, 12/2022, A5E03314905-AB