Electric connections

Unsafe connection
Do only use sensor connection cable(s) and hybrid cable(s) delivered by Siemens.
4.1

Safety notes

4.1.1

Unsuitable cables, cable glands and/or plugs

Unsuitable cables, cable glands and/or plugs
Risk of explosion in hazardous areas.
• Use only cable glands/plugs that comply with the requirements for the relevant type of
• Tighten the cable glands in accordance with the torques specified in Technical data
• Close unused cable inlets for the electrical connections.
• When replacing cable glands, only use cable glands of the same type.
• After installation, check that the cables are seated firmly.
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
• For the electrical connection specifications, refer to the information in Technical data
• At the mounting location of the device observe the applicable directives and laws for
LDS 6 & Optics housing CD 6 Compact Operating Instructions
Compact Operating Instructions, 12/2022, A5E03314905-AB
WARNING
WARNING
protection.
(Page 59).
WARNING
(Page 59).
installation of electrical power installations with rated voltages below 1000 V.
4
31