Turn the device on & off
Line Integrity Test
Honeywell offers the optional capability to check for leaks in Midas S2
sample lines. Midas S2 Transmitter for LIT (line Integrity Test) detects any
leak in the tubing with a pneumatic signal change from the valves
installed at the end of the sample line. This test is performed automatically
as a LIT. The Line Integrity Test Option requires the Calibration process.
If you make any changes to the Pump Module, Check Valve,
l
Filter, and Tube, and start a new LIT calibration again.
Verify the Check Valve flow direction during installation.
l
Verify both the Check Valve and Tube are firmly assembled.
l
Mark the insertion from the end of the tube to the length of
l
15.5mm. When inserting the tube, make sure the insertion is up
to the marked position.
The external filter is installed further from the Transmitter, and
l
the Check Valve is installed closer to the transmitter.
CAUTION
RISK OF MALFUNCTION
17