9. GUIDANCE AND MANUFACTURER'S DECLARATION
Guidance and manufacturer's declaration - electromagnetic immunity
OMRON BPM is intended for use in the electromagnetic environment specified below. The customer
or the user of this OMRON BPM should assure that it is used in such environment.
IEC 60601
Immunity test
Conducted RF
3 V rms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Compliance
test level
level
Not
Applicable.
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the OMRON
BPM including cables, than the recommended
separation distance calculated from the equation
appropriate to the frequency of the transmitter.
Recommend separation distance
Not Applicable.
d = 1.2
P
80 MHz to 800 MHz
d = 2.3
P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey,
should be less than the compliance level in each
b
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol: