Specifications; Electrical Specifications; Refrigeration Specifications; Installation Specifications - Follett K25541 Installation, Operation And Service Manual

Undercounter refrigerators
Table of Contents

Advertisement

Specifications

REF4P
31.38" (79.7 cm) H x 23.75"
(60.3 cm) W x 27" (68.5 cm) D
REF5P
34.00" height (86.4 cm) H x 23.75"
(60.3 cm) W x 27" (68.5 cm) D
REF5BBP-T
34.00" (86.3 cm) H x 23.75"
(60.3 cm) W x 27" (68.5 cm) D

Electrical Specifications

§ 115 V, 60 Hz, 1 phase
§ Full load: 4.1A
§ Minimum circuit ampacity: 15A
§ Connect to dedicated circuit, fuse or breaker
§ Maximum size of branch circuit overcurrent device: 15A
§ Follett recommends circuit be protected by GFCI

Refrigeration Specifications

Refrigerant
R134A

Installation Specifications

Ambient temperature must not exceed 39 C (100 F).
The front louvered panel must be kept free of any cabinet trim or obstructions to ensure proper ventilation of the
refrigeration system.
CAUTION!
§ Equipment must be wired according to local and national electrical codes.
§ Always disconnect power before servicing refrigerator.
4
REF4P, REF5P and REF5BBP-T Undercounter Refrigerators
Fits below 34" (86.4 cm) high ADA-
compatible counter
Fits below standard 36" (91.4 cm)
high counter
Fits below standard 36" (91.4 cm)
high counter
Charge Size (oz)
9
Maximum Design Pressures (psi)
High Side
Low Side
190
3.9 cu ft capacity
4.5 cu ft capacity
2 drawers with total storage of 40
cardboard plasma packs 300 ml/
pack); max. drawer load of 35 lb
(16 kg) each
110