v3001 FC
Calibration Manual
Password Entry
To go to the calibration mode, push five times until CAL is shown. You will
need to enter a password to access calibration mode.
To enter the password:
1. Push and the CAL counter is shown. For example n002.
2. Push to show "????".
3. Push to change the first digit of your password (default: 1234).
4. Push to confirm your choice. The subsequent "?" changes to "_".
5. Do steps 3 and 4 again to enter the subsequent digits of the four-number
6. When all of the correct digits are entered, push to confirm the input.
Change the Password
If you change the password and then lose it, see the "Restore the
Default Password" section.
To change the password:
1. Do steps 1 through 5 in the "Password Entry" section.
2. Before you push to confirm your final input (step 6), push to show "----"
3. Push to change the first "-" to the first digit of your new password.
4. Push to confirm your choice.
5. Repeat steps 3 and 4 to enter the subsequent digits of the new four-number
6. When the correct digits are entered, push to confirm the input and change
8
password.
If the correct password is entered, "C-01" is shown. If the incorrect password
is entered, "????" is shown and the password must be correctly entered.
on the display.
password.
the password. If the Product has been calibrated, it will go to normal
measurement mode.
Note