Compressor Troubleshooting - Parker Balston Super Flow Tri-Gas Generator Series Installation, Operation And Maintenance Manual

Table of Contents

Advertisement

Bulletin TI-LCMS-SF5000J
1-800-343-4048
Troubleshooting
Disconnect the electrical power and depressurize the generator before attempting any trouble-
shooting activities according to the shut down steps in the operation section. Only trained per-
sonnel using reasonable care should perform any troubleshooting activities.
Determine whether the trouble is with the compressed air source or the Tri-Gas Generator. First,
check that the compressed air source is delivering. If you have an integrated compressor (Model
LCMS-SF5000), refer to the compressor troubleshooting section, otherwise check that the
compressed air source is delivering a minimum, continuous flow of 180 SLPM (6.3 SCFM) at a
pressure between 110 - 140 PSIG (7.6 - 9.6 barg).
If the compressor air supply is confirmed, then go to the Tri-Gas Generator troubleshooting sec-
tion, and begin troubleshooting the Tri-Gas Generator.

Compressor Troubleshooting

Symptom
POWER ON light (green)
does not appear
POWER ON light is on but
motor will not start
Compressor is running but
will not make pressure
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
Probable Cause
• Power not plugged in sup-
ply outlet.
• Main disconnect is not ON.
• Blown fuse or tripped OFF
circuit breaker at customer
provided power supply.
• Blown fuse at primary side
of transformer.
• LED power switch has
failed.
• Motor overload has tripped
• Wrong or low voltage
• Motor failed
• Depressurize Valve on
TriGas Generator is OPEN
• Auto Drain Valve is open
continuously
• Drive belts broke or are too
loose
• Clogged intake filter ele-
ment
• Pressure relief valve has
opened
• Excessive tip seal wear
• Motor running wrong
direction
30
Corrective Action
• Plug in power cord.
• Switch disconnect to ON.
• Replace fuse or switch
disconnect to ON.
• Replace fuse on primary
side. Be sure to use the
same type and size.
• Replace switch.
• Determine fault and reset
overload (Blue button)
• Check incoming power sup-
ply and unit power rating
• Replace motor
• Close Valve
• Replace Auto Drain Valve
• Replace drive belts and (or)
tighten to specification
• Replace intake filter element
per instructions
• Pressure switch needs
replacing or motor contacts
welded shut
• Replace tip seals per in-
structions every 5,000 hrs
• Correct power connections
www.labgasgenerators.com