The compressor drain tube is a black nylon 1/4" tubing connecting the drain outlet to the provided
disposal container (see figure 12). A blue silencer is provided to reduce the discharge noise.
The Tri-Gas generator drain tube is also a black nylon 1/4" tubing connecting the drain outlet to
the same provided disposal container (see figure 13).
The three outlet gas fittings are 1/4" stainless steel compression type (see figure 14) . Piping to
the instrument should be PFA plastic, clean stainless steel or pre-cleaned refrigeration grade cop-
per with a minimum pressure rating of 125 psig (8.6 bar). Do not use any liquid, thread sealing
compounds which could contaminate the gases. Use only PTFE sealant tape. Check all piping
connections for leaks using a bubble-type leak detecting solution. Europe models include tube
end converters to adapt to 6mm tubing.
Figure 13 Tri-Gas Drain to Disposal
Container (One Tri-Gas Generator
shown)
A03-0286 voltage step-down transformer - If your facility supply voltage is 255V or higher,
then the use of this transformer is required.
A03-0281 Voltage Step-up Transformer - If the supply voltage is 195V or lower, then the use
of this transformer is required.
Model 76080 halogenated hydrocarbon scrubber for model LCMS-5001 NT: If the com-
pressed air supply to the generator is subject to contamination from halogenated hydrocarbons,
install a halogenated hydrocarbon scrubber directly upstream from the generator. Exposure of
the generator to halogenated hydrocarbons will damage the catalyst module and void the
warranty. For additional information, request the Analytical Gas Supplies Product Catalog from
your local representative.
IKLCMS-5000: Installation kit includes PFA tubing, tube cutter, PTFE sealant tape.
IKLCMS-5000UK: Installation kit for Europe includes same items as above with the addition of
6mm tubing and 6mm tube end adapters.
LCMS-MANIFOLD: Manifold consolidating six outlets to three outlets (see Figure 51 and Bulletin
TI-B06-0092 for details).
LCMS-SFEZLINK: Gas back-up panel for cylinder reserve: If there is a loss of electrical power or
pressure, the gas back-up panel will automatically switch over to a single or bank of nitrogen gas
cylinders (see Figure 52 and Bulletin TI-B06-0092 for details).
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
Figure 14 Three Outlet Gas Fittings
(One Tri-Gas Generator shown)
10
www.labgasgenerators.com

Advertisement

Table of Contents
loading

Related Products for Parker Balston Super Flow Tri-Gas Generator Series

This manual is also suitable for:

Lcms-sf5000Lcms-sf5001tLcms-sf5001nt

Table of Contents