System Specifications

System Specifications
CSA Certification Standard
IEC Standard
CENELEC Standard
UL Standard
IEC 61010 Installation
IEC 61010 Pollution
EMC Compliance
Curtain Gas (N2) Purity
Curtain Gas Max. Flow/Pressure
Max. Inlet Hydrocarbon Content
Source Gas (UHP Zero Grade Air) Purity
Source Gas Max. Flow/Pressure
Exhaust Gas (Dry Air) Purity
Exhaust Gas Max. Flow/Pressure
Suspended Liquids
Particles > 0.01 Micron
Phthalates
Sound Level
Min./Max. Operating Pressure
Flow Capacity
Inlet Air Consumption
Inlet Air Temp. Range
Min./Max. Ambient Temp.
Max. Ambient Relative Humidity
Altitude
Heat Dissipation - Compressor
Heat Dissipation - Tri-Gas Generator
Electrical Requirements
- Compressor 60Hz
1
Electrical Requirements
- Compressor 50Hz
1
Electrical Requirements
- Gas Generator
1
Power Consumption - Compressor
Power Consumption - Tri-Gas Generator
Start-Up Time
Compressor Outlet - Compressed Air
Compressor Outlet - Tank Drain
Tri-Gas Generator Ports - Comp Air Inlet
Tri-Gas Generator Ports - Drain
Tri-Gas Generator Ports - Gas Outlets (3)
Compressor Fuses - Primary FU1, FU2
Compressor Fuses - Secondary FU3
Tri-Gas Fuses - 120V
Tri-Gas Fuses - 230V
Dimensions LCMS-SF5000
Dimensions LCMS-SF5001T
Dimensions LCMS-SF5001NT
Product Weight LCMS-SF5000
Product Weight LCMS-SF5001T
Product Weight LCMS-SF5001NT
Shipping Weight LCMS-SF5000
Shipping Weight LCMS-SF5001T
Shipping Weight LCMS-SF5001NT
Notes:
1. Main supply voltage fluctuations not to exceed ± 10% of nominal voltage.
2. Optimum minimum inlet pressure is 110 psig (7.6 barg) to achieve published gas specifications of instrument manufacturers.
3. Consult factory for performance specifications at lower inlet pressures, or high elevation installations.
1-800-343-4048
4. PCS motors installed after 11/3/09, starting with Serial Number LCMS-SF5000NA-1010.
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
LCMS-SF5000
CAN/CSA C22.2 No. 61010-1
IEC 61010-1: 2001
EN 61010-1: 2001
UL 61010A-1: 2002
Category II
Degree 2
EN61326-1: 1997/CISPR 11: 1990/IEC 1000
99.999% Organics, 99% Oxygen
12 SLPM/80 PSIG (5.5 Barg)
100 PPM
<0.1 PPM Total Hydrocarbons
40 SLPM/110 PSIg (7.6 Barg)
-40°F (-40°C) ATM Dew Point
16 SLPM/60 PSIG (4.1 Barg)
None
None
None
< 49 dB(A)
3
110/140 PSIG (7.6/9.6 Barg)
See Pressure/Flow Charts, page 28
180 SLPM (6.3 scfm)
60°F-95°F (15°C-35°C)
60°F/90°F (16°C/32°C)
80%
2000M
10246 BTU/Hour
1229 BTU/Hour
195-254VAC, 60Hz, 1 Phase, 14A
230VAC, 50Hz, 1 Phase, 13A
100-120/200-240VAC, 60/50Hz, 1 Phase, 4A/2A
13A@230VAC, 3000 Watts
3A@120VAC, 360 Watts
1.5A@230Vac, 360 Watts
45 Minutes
3/8" Tube
1/4" Tube
3/8" Tube / 3/8" BSPT Adapter
1/4" Tube
1/4" SS Compression / 6 mm adapters
10.3 x 38.1mm, Type T, 600V, 1.25A
6.3 x 31.7mm, Type T, 250V, 1.25A
5 x 22mm, Type T, 250V, 4A
5 x 22mm, Type T, 250V, 2A
34"L x 61"W x 43"H
(86cm x 155cm 109cm)
-------
-------
788 lbs (357 kgs)
-------
-------
1166 lbs (529 kgs)
-------
-------
25
LCMS-SF5001T, LCMS-SF5001NT
same
same
same
same
same
same
same
See Note 2
same
same
See Note 2
See Note 2
See Note 2
same
same
same
same
< 30dB(A)
85/140 psig (5.8/6.9 barg)
same
same
same
same
same
same
-------
same
4
-------
-------
-------
-------
same
same
same
-------
-------
same
same
same
-------
-------
same
same
-------
61"L x 34"W x 43"H
(155cm x 86cm x 109cm)
21"L x 46"W x 41"H
(53cm x 116cm x 104cm)
-------
448 lbs (203 kgs)
220 lbs (100 kgs)
-------
688 lbs (312 kgs)
450 lbs (204 kgs)
www.labgasgenerators.com