Fuse Replacement - Parker Balston Super Flow Tri-Gas Generator Series Installation, Operation And Maintenance Manual

Table of Contents

Advertisement

Bulletin TI-LCMS-SF5000J
Figure 45 Inlet / Outlet
Figure 47 Nut

Fuse Replacement

1-800-343-4048
Figure 48 Disconnect Electrical
The Tri-Gas Generator fuses are located in the power receptacle on the top, rear side of the gen-
erator. Before servicing the fuses, turn the generator off and disconnect the power cord
from both the power supply and the generator power receptacle.
To access the fuses, use a small screwdriver to remove the fuse holder located in the power re-
ceptacle of the generator (see figure 50). Insert screwdriver blade in holder tab and push in and
draw out. In the generator, both the phase and neutral are fused separately. As a result of this
configuration, both fuses should be checked any time fuse replacement is warranted. Replace
either one or both fuses as necessary and re-assemble. To maintain the safety and perfor-
mance integrity of the product, use only fuses of the size and type detailed in the Specifi-
cations section of this manual.
Figure 50 Tri-Gas Fuse
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
Figure 46 Dryer Membrane
24
Figure 49 Disconnect Electrical at PCB
www.labgasgenerators.com