Visual Inspection, Cleaning And Disinfection Frequency - GE MAC 5 A4 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Cleaning and Disinfection
• Avoid exposure to highly alkaline conditions (pH > 11), as this can damage
products (for example, aluminum parts).
• Never use conductive solutions or solutions that contain wax or wax compounds
to clean the equipment.
• Do not immerse the device in any liquid as this may corrode metal contacts and
affect signal quality.
• Do not drip or expose the writer assembly to any liquids.
• Do not allow fluid to pool around connection pins. If this happens, blot dry with a
soft, lint-free cloth.
• Avoid contact with open vents, plugs, or connectors during the cleaning and
disinfecting procedures.
• Never autoclave or steam-clean the device.
• Do not use until thoroughly dry.
• Do not use any of below materials to clean the device, because their use may
damage equipment surfaces.
• Organic solvents
• Abrasive cleaners or solvents of any kind
• Acetone
• Ketone
• Betadine
• Sodium salts

Visual Inspection, Cleaning and Disinfection Frequency

The table indicates the frequency of visual inspection, cleaning, and disinfection
procedures.
Component
Device and Trolley
NOTE:
Trolley is an
optional purchase.
Leadwires
Reusable electrodes
270
Visual Inspection
Daily, preferably before
the equipment's first
use each day
Refer to the supplier's instructions for leadwire cleaning and disinfection.
Refer to the supplier's instructions for reusable electrode cleaning,
disinfection and sterilization.
  MAC
5 A4/MAC
5 A5/MAC
Cleaning
Monthly, or more
frequently, as needed
5 Lite
Disinfection
Follow the same
frequency as cleaning.
Disinfection must
be performed after
cleaning.
 5864335-001-1