Preventative Maintenance; General; Integrity Of Hoses And Cuffs; Visual Inspection - GE DINAMAP ProCare Service Manual

Table of Contents

Advertisement

Preventative Maintenance

General

Integrity of Hoses and Cuffs

Visual Inspection

Revision C
Maintenance: Preventative Maintenance
WARNING
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this device, to implement the
recommended maintenance schedule may cause equipment
failure and possible health hazards. The manufacturer does
not, in any manner, assume the responsibility for performing
the recommended maintenance schedule, unless an
Equipment Maintenance Agreement exists. The sole
responsibility rests with the individuals, hospitals, or institutions
utilizing the device.
Preventative maintenance tasks include cleaning the ProCare Monitor, checking
pressure calibration, pneumatic leakage, pneumatic system overpressure point,
temperature calibration (200 and 400), and verification of the SpO
and 400). Perform the following maintenance procedures as required.
When the pneumatic integrity of any BP cuff and hose is in doubt, replace the
cuff and hose, and discard the questionable accessories.
The monitor and its components should be carefully inspected prior to
installation, once every 12 months thereafter and each time the equipment is
serviced.
Carefully inspect the equipment for physical damage to the case, the display
screen, and the keypad. Do not use the monitor if damage is determined.
Refer damaged equipment to qualified service personnel.
Inspect all external connections for loose connectors or frayed cables.
Have any damaged connectors or cables replaced by qualified service
personnel.
Inspect the display face for marks, scratches, or other damage.
Physical damage to a flat panel display glass may pose an implosion
hazard. Have the flat panel display replaced by qualified service personnel if
necessary.
Safety labels and inscription on the device are clearly legible.
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
system (300
2
4-3

Hide quick links: