Ranger

Introduction

The 3M
warming set. The Ranger blood/fluid warming system is designed to warm blood, blood products,
and liquids and deliver these at flow rates from KVO to 500 mL/min. At these flow rates, the
device maintains fluid output temperatures ranging from 33°C to 41°C (Note: Output temperatures
are dependant on input fluid temperatures and flow rate). It takes less than 2 minutes to warm up
to the 41°C set point temperature.
Disposable blood/fluid warming sets are available in: pediatric flow, standard flow, and high flow
applications. Warming sets are sterile, latex-free, single-use-only items and are designed to be used
with the warming unit.
The Ranger blood/fluid warming unit is designed to be mounted to an I.V. pole. A handle located
on the top of the unit makes transport easy. When mounted to the I.V. pole, the unit fits easily
above the 3M
fluid warming sets visit us online at rangerfluidwarming.com.
This manual includes operating instructions and unit specifications for the Ranger blood/fluid
warming system. For information about using Ranger blood/fluid warming sets with the Ranger
blood/fluid warming unit, Model 245, refer to the "Instructions for Use" included with each
warming set. The Ranger blood/fluid warming system should only be used in healthcare facilities by
trained medical professionals.

Indications for use

The Ranger blood/fluid warming system is intended to warm blood, blood products, and liquids.

Definition of Symbols

English 202457B
Blood/Fluid Warming System
Ranger
blood/fluid warming system includes a warming unit and a disposable
TM
TM
Bair Hugger
warming unit. For additional information on the Ranger blood/
TM
TM
ON
OFF
Fuse
CAUTION
Type B Applied Part
Voltage, Alternating Current (AC)
An equipotentiality plug (grounded) conductor other than a protective earth
conductor or a neutral conductor, providing a direct connection between electrical
equipment and the potential equalization busbar of the electrical installation.
Please consult IEC 6060-1; 2005 for requirements.
www.rangerfluidwarming.com reorder #202457B
English 3