Specifications - 3M Ranger 245 Operator's Manual

Blood/fluid warming system
Hide thumbs Also See for Ranger 245:
Table of Contents

Advertisement

Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
16 English

Specifications

Physical characteristics
WARMING UNIT
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)
CERTIFICATIONS
IEC/EN 60601-1; CAN/CSA-C22.2, No.601.1
CLASSIFICATION
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2,
No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No.4HZ8
Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I,
Type B, Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with
respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN 60601-1 and
in accordance with Canadian/CSA C22.2, No. 601.1. Classified under the Medical Device Directive
as a Class IIb device.
Electrical characteristics
DEVICE RATING
100-120 VAC, 50/60 Hz
220-240 VAC, 50/60 Hz
MAXIMUM HEATING POWER
900 W
FUSE
2 x T10A-H (250V) for 100-120 VAC
2 x T6.3A-H (250V) for 220-240 VAC
FUSE TYPE
Time delay, high breaking
Leakage current
Meets leakage current requirements in accordance with UL/IEC 60601-1.
Environmental conditions
OPERATING TEMPERATURE RANGE
15° to 40°C (59° to 104°F)
STORAGE TEMPERATURE RANGE
-20° to 45°C (-4° to 113°F)
OPERATING HUMIDITY
10 to 85% RH, noncondensing
ATMOSPHERIC PRESSURE RANGE
50 kPa to 106 kPa
3M
Ranger
Blood/Fluid Warming System
Temperature characteristics
SET POINT TEMPERATURE
41°C
-
OVER
TEMPERATURE ALERT
43°C
-
UNDER
TEMPERATURE ALERT
33°C
-
OVER
TEMPERATURE CUTOFF
44°C (warming units Rev N and newer)
46°C (warming units Rev A to M)
www.rangerfluidwarming.com reorder #202457B