3M Ranger 245 Operator's Manual page 19

Blood/fluid warming system
Hide thumbs Also See for Ranger 245:
Table of Contents

Advertisement

Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
3M
Ranger
Method
1. Unplug the warming unit.
2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive
solution, such as Alconox
3. Insert the tool from the back of the unit and pull the tool all the
way out from the front.
4. Rinse the tool with water and repeat 3 times.
5. Wipe off the unit to remove excess fluid.
TO CLEAN RESISTANT
1. Spray a nonabrasive solution inside the slot of the warming unit and
let sit for 15-20 minutes.
2. Clean the unit by using the cleaning tool.
Note: If the cleaning tool cannot be inserted through the slot of the
warming unit because of excess dried-on fluids, send the unit to a
biomedical technician.
Storage
Store all components in a cool, dry place when not in use.
Servicing
There are no user-serviceable parts in the Ranger blood/fluid warming unit. All service must be
performed by 3M Patient Warming or an authorized service technician. Call 3M Patient Warming
Technical Service at 800-733-7775 or 952-947-1200 for service information. Outside of the USA
contact your local 3M Patient Warming representative.
Alconox is a trademark of Alconox, Inc.
English 202457B
Blood/Fluid Warming System
brand detergent.
TM
,
-
:
DRIED
ON FLUIDS
www.rangerfluidwarming.com reorder #202457B
English 15