Remote Alarm Terminal; Air Intake Port - Panasonic MDF-U76VA Operating Instructions Manual

Ultra-low temperature freezer
Hide thumbs Also See for MDF-U76VA:
Table of Contents

Advertisement

REMOTE ALARM TERMINAL

The terminal of the remote alarm is installed at the lower left side of the unit. The alarm is outputted from
this terminal. Contact capacity is DC 30 V, 2 A.
Contact output:
At normal
At abnormal
Note:
• The buzzer is silenced by pressing buzzer stop key (BUZZER) on the control panel during alarm
condition. (A remote alarm is continuing the operation.)
The buzzer will be activated again after certain suspension if the alarm condition is continued.
• The alarm is actuated when the power supply plug is disconnected from the outlet or the power switch is
OFF.

AIR INTAKE PORT

When a door is closed and opened soon, a door does not
sometimes open this product.
The warm air which went into the chamber is cooled down rapidly,
and this is because air inside the chamber contracted.
When a door is closed and opened soon, a door is easy to open
with the following process.
1. Turn the cap of the left side counterclockwise about two laps.
(That can be removed completely)
2. Put the air into the chamber from about twenty seconds, and
open a door.
3. Close a cap if a door opens.
A door may not open in the above method when there are frost and ice in the air intake port. In that case,
open a cap, and investigate the matter whether there is no frost inside the air intake port. Remove frost
inside the air intake port in "the air intake port cleaning stick" of the accessories when there is frost.
Clean a air intake port every month even when there is no frost inside the air intake port. Refer to page
24.
WARNING
For removing the frost of the air intake port, do not use a tool with sharp edge such as a knife or a screw
driver.
Replace the cap when the air intake port is not used. Improper replacement may cause rise of chamber
temperature or condensation around the air intake port.
between COM. and N.O.
Open
Close
Use a twisted sealed wire for the connection.
Type UL2343, UL2448, UL2464, UL2552, UL 2623
Length: 30 m max.
between COM. and N.C.
Close
Open
11
Two laps
Cap