Cleaning and disinfection
– The temperature of the first water inflow must be below
– The program does not include any sudden changes in
– The program must at least include the following steps:
Olympus has validated automatic cleaning with the following
program:
– Vario TD program
– Pre-cleaning for 4 min.
– Cleaning for 10 min. at 55 °C (131 °F)
– Neutralization for 6 min.
– Final rinsing for 3 min.
Removing from washer-disinfector
1. Remove the product from the washer-disinfector.
2. Let the product cool down at room temperature.
3. If the product is wet, dry the product.
8.5 Inspection
This section describes all necessary steps to make the product fit
for use again:
Checking for mechanical defects
1. Check that the product has:
46
25 °C (77 °F) to avoid protein fixation.
temperature.
– Pre-cleaning
– Cleaning
– Neutralization
– Final rinsing with deionized water
To support drying, use a clean lint-free cloth or filtered
compressed air.
– no dents, cracks, kinks, or deformations
– no cuts and other defects on the outer sleeve of the cable
– no scratches
– no corrosion
– no missing or loose parts
– no cuts and other defects on the insulation