4.3 Connecting the HF cable (not for WA00014A)

CAUTION
Risk of injury to the patient and the user
Accidentally disconnecting the HF cable can cause sparkover
leading to electrical injury, thermal injury or unintended nerve
stimulation.
● Check that the HF cable is securely connected to the
NOTICE
Risk of damage to the product
Connecting the HF cable to the wrong socket of the
electrosurgical generator can damage the HF cable and the
electrosurgical generator.
● Make sure to connect the HF cable to correct and intended
1. Connect the HF cable to the compatible instrument.
2. Connect the HF cable to the compatible electrosurgical

4.4 Connecting the HF cable (only for WA00014A)

CAUTION
Risk of injury to the patient and the user
Accidentally disconnecting the HF cable can cause sparkover
leading to electrical injury, thermal injury or unintended nerve
stimulation.
instrument and the electrosurgical generator.
socket on the electrosurgrical generator.
generator.
Refer to the instructions for use of the electrosurgical
generator and the instrument.
Before use
25