Purging - Siemens LDS 6 Operating Instructions Manual

In-situ laser gas analyzer
Table of Contents

Advertisement

3.5

Purging

Purging can be divided into purging of optical process interface and purging of sensors.
Sensor Head Purging
Ambient air contains roughly 1% water vapor. Therefore the sensor heads must be purged
with a very dry gas like nitrogen. Since the resolution of the water vapor application of LDS 6
is better than 1 ppm the nitrogen should be correspondingly dry.
The recommended flow is 10 l/min and the absolute minimum is 6 l/min. The pressure
should be as close to ambient as possible and should not exceed 1.5 bar absolute. Too low
flow rate and/or to humid purging gas might lead to faulty measurements.
Figure 3-18
1. Sensor side purging input.
2. Purging bypass tube.
3. Lens.
4. Sensor side purging output check valve.
Sensor CD 6C Operating Instructions
Operating Instructions, 12/2007, A5E00912721-02
Gas purging in sensor head
4
2
3
Installation Guidelines
3.5 Purging
1
25