Purging can be divided into purging of optical process interface and purging of sensors.
Sensor Head Purging
Ambient air contains roughly 1% water vapor. Therefore the sensor heads must be purged
with a very dry gas like nitrogen. Since the resolution of the water vapor application of LDS 6
is better than 1 ppm the nitrogen should be correspondingly dry.
The recommended flow is 10 l/min and the absolute minimum is 6 l/min. The pressure
should be as close to ambient as possible and should not exceed 1.5 bar absolute. Too low
flow rate and/or to humid purging gas might lead to faulty measurements.
Figure 3-18
1. Sensor side purging input.
2. Purging bypass tube.
3. Lens.
4. Sensor side purging output check valve.
Sensor CD 6C Operating Instructions
Operating Instructions, 12/2007, A5E00912721-02
Gas purging in sensor head
4
2
3
Installation Guidelines
3.5 Purging
1
25