5-3. How to Make Measurements
_
WARNING
1. To avoid electrical shock hazard and/or damage to the
meter, do not measure voltages that might exceed 600V
above earth ground.
2. Before using the instrument, inspect the test leads,
connectors, and probes for cracks, breaks, or grazes in
the insulation.
DC Voltage Measurement
1. Plug the red test lead into the "V Ω mA" jack and the
black test lead into the "COM" jack.
2. Set the RANGE switch to the desired DCV position. If
the voltage to be measured is not known beforehand,
set the switch to the highest range and reduce it until a
satisfactory reading is obtained.
3. Connect the test leads to the device or circuit being
measured.
4. Turn on the power of the device or circuit being
measured. The voltage value will appear on the digital
display along with the voltage polarity.
-10-
Resistance
Range
Resolution
200Ω
0.1Ω
2kΩ
20kΩ
10Ω
200kΩ
100Ω
2000kΩ
1kΩ
OVERLOAD PROTECTION: 15 seconds maximum 220Vrms
on all ranges.
MAXIMUM OPEN CIRCUIT VOLTAGE: 2.8V.
Audible Continuity
Range
Description
Built-in buzzer sounds if resistance is
less than 100Ω.
O VERLOAD PROTECTION: 220Vrms for a maximum of 15
seconds.
TR h
FE
R ange
Base Current
NPN
@2.8V DC
PNP
@2.8V DC
DIODE TEST
Measures forward resistance of a semiconductor junction
in kΩ at max. test current of 1.0mA.
Accuracy
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(1% of rdg+2dgt)
Test Condition
10μA DC
10μA DC
-7-