Troubleshooting - Braun ThermoScan 4 Manual

Ear thermometer
Hide thumbs Also See for ThermoScan 4:
Table of Contents

Advertisement

Available languages

Available languages

Troubleshooting

Situation
No hygienic lens
filter is attached .
Measurement
before device
stabilization .
Ambient
temperature is not
within the allowed
operating range
( 50–104 °F or
10–40 °C) .
HI = too high
LO = too low
Temperature taken
is not within typical
human temperature
range (93 .2–108 °F
or 34–42 .2 °C ) .
System error .
or
Battery is low .
Battery is too low .
Device cannot be
powered on to the
or
ready state .
10
Solution
Attach new, clean, hygienic
lens filter .
Wait until all the icons stop
flashing .
Allow the thermometer to remain
for 30 minutes in a room where
the temperature is between
50 and 104 °F or 10 and 40 °C .
Make sure the probe tip and lens
are clean and a new, clean,
disposable lens filter is attached .
Make sure the thermometer is
properly inserted . Then, take a
new temperature .
Reset the thermometer
by removing the battery and then
replacing it again, wait 1 minute,
then turn on again .
Please contact Consumer
Relations (See Guarantee Card) .
Insert new battery .
Insert new battery .