Attaching the area dose measurement device
Hide thumbs Also See for SIREMOBIL Compact:

Advertisement

6
Programming and adjusting the area dose measurement device
NOTE
Calculating the CONSTANT parameter (Qt * Kg)
Read and record the Kg value shown on the electronics unit of the area dose
measurement device.
Read the Qt value on the area dose chamber of the area dose measurement device.
Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Parameter menu.
Select "Diamentor" in the combo box "Parameter Groups:".
Click "Get from Unit".
Select "1. Diamentor Existing".
Select 'yes' in the "Value actual" field.
Click "Set Value".
Select "2. Constant (Qt * Kg)" .
Enter the parameter you have calculated in the "Value Actual:" field.
Click "Set Value".
Click "Put to Unit".
Click "Save to File".
SIREMOBIL Compact
Installation and Setting Instructions
This adjustment is used to calibrate the area dose measurement
device of the SIREMOBIL Compact display.
Register 3
RXR2-130.033.01
Rev. 04
Page 6 of 10