Programming and adjusting the area dose measurement device
NOTE
Calculating the CONSTANT parameter (Qt * Kg)
•
Read and record the Kg value shown on the electronics unit of the area dose
measurement device.
•
Read the Qt value on the area dose chamber of the area dose measurement device.
•
Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming
•
Connect the service PC to the serial interface on board D1.
•
Start the service program.
•
Select the Adjustment, Parameter menu.
•
Select "Diamentor" in the combo box "Parameter Groups:".
•
Click "Get from Unit".
•
Select "1. Diamentor Existing".
•
Select 'yes' in the "Value actual" field.
•
Click "Set Value".
•
Select "2. Constant (Qt * Kg)" .
•
Enter the parameter you have calculated in the "Value Actual:" field.
•
Click "Set Value".
•
Click "Put to Unit".
•
Click "Save to File".
SIREMOBIL Compact
Installation and Setting Instructions
This adjustment is used to calibrate the area dose measurement
device of the SIREMOBIL Compact display.