Fluke 287C Calibration Manual page 30

True-rms digital multimeters
Table of Contents

Advertisement

287/289
Calibration Manual
Rotary Switch
Step
30
mV DC
Ω
31
Ω
32
33
Ω
Ω
34
Ω
35
Ω
36
Ω
37
38
Ω
39
Diode Test
40
Diode Test
41
μA DC
μA DC
42
μA AC
43
44
μA AC
45
mA DC
46
mA AC
47
mA AC
48
mA AC
49
ADC
50
ADC
51
AAC
52
AAC
26
Table 7. Calibration Adjustment Steps for Firmware Version 1.00 (cont.)
Source Value
Position
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
500.0 mV, 60.0 Hz
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
0 Ω, zcomp wire2
500 Ω, zcomp wire2
5 kΩ, zcomp wire2
50 kΩ, zcomp wire2
Disconnect 5520A 4-Wire configuration and turn COMP off
500 kΩ
5.0 MΩ
50.0 MΩ
500.0 MΩ
0 Ω, zcomp wire2
3.1 V, 0 Hz
500.0 μA, 0 Hz
5000.0 μA, 0 Hz
500.0 μA, 60 Hz
5000.0 μA, 60 Hz
50.0 mA, 0 Hz
400.0 mA, 0 Hz
50.0 mA, 60 Hz
400.0 mA, 60 Hz
5.0 A, 0 Hz
10.0 A, 0 Hz
5.0 A, 60 Hz
10.0 A, 60 Hz
Settling Time
Calibrating Time
(seconds)
5
4
5
5
5
5
5
5
5
2
2
3
3
10
10
5
3
10
10
8
3
10
9
[1]
(seconds)
3
20
2
2
2
5
5
5
2
2
2
4
4
4
3
4
4
3
4
4
2
4
4