Fluke 287C Calibration Manual page 29

True-rms digital multimeters
Table of Contents

Advertisement

Table 7. Calibration Adjustment Steps for Firmware Version 1
Rotary Switch
Step
Position
1
mV DC
2
mV DC
3
mV DC
4
mV DC
5
mV AC
6
mV AC
7
mV AC
8
mV AC
9
mV AC
10
VAC
11
VAC
12
VAC
13
VAC
14
VAC
15
VAC
16
VAC
17
VAC
18
VDC
19
VDC
20
VDC
21
VDC
22
VDC
23
VDC
24
VDC
25
VDC
26
VDC
27
VDC
28
VDC
29
mV DC
Source Value
0.0 V, 0 Hz
500.0 mV, 0 Hz
50.0 mV, 0 Hz
50.0 V, 0 Hz
5.0 mV, 60 Hz
50.0 mV, 60 Hz
500.0 mV, 60 Hz
50.0 mV, 60 Hz
0.0 V, 0 Hz
0.0 V, 0 Hz
5.0000 V, 60 Hz
[2]
5.0000 V, 65 kHz
50.000 V, 60 Hz
50.000 V, 100 kHz
250.00 V, 60 Hz
500.00 V, 60 Hz
1000.0 V, 60 Hz
0.0 V, 0 Hz
5.0000 V, 0 Hz
5.0 V, 60.0 Hz
5.0 V, 20.0 kHz
50.000 V, 0 Hz
50.0 V, 60.0 Hz
50.0 V, 20.0 kHz
500.00 V, 0 Hz
500.0 V, 60.0 Hz
1000.0 V, 0 Hz
1000.0 V, 60.0 Hz
50.0 mV, 60.0 Hz
True-rms Digital Multimeters
Settling Time
Calibrating Time
(seconds)
2
2
2
7
4
4
4
10
10
4
10
5
6
6
8
6
7
3
3
4
3
3
5
7
3
4
2
4
4
Calibration Adjustment
[1]
(seconds)
10
15
2
15
4
2
5
5
5
10
5
35
10
120
2
5
4
30
5
5
35
8
15
115
7
5
5
5
3
25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

289287287/fvf289/imsk287/eur

Table of Contents