To Assemble the EI Ion Source

Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
Always wear clean gloves when working in the analyzer chamber to avoid
C A U T I O N
contamination.
1
2
3
4
5
6
7
8
While installing, do not overtighten the repeller nut or the ceramic repeller insulators
C A U T I O N
will break when the source heats up. The nut should only be finger-tight.
7200 Accurate-Mass QTOF Operation Manual
Large (8650-0030)
Small (8650-0029)
Assemble the repeller assembly, by attaching the repeller to the repeller
cap and connecting the clocking button with the screw.
Align the heater sensor with the source body.
Slide the repeller cap onto the heater sensor, and place the ceramic
repeller insulator on top, aligning the screw holes.
Place the RIS bayonet mount on top of the ceramic repeller insulator, and
attach with the washers and screws.
Attach the extractor lens insulator to the extractor lens, and slide them
onto the other end of the source body
Place the ion focus insulator on the ion focus lens and attach with the two
screws.
Set the ion focus assembly on top of the extractor lens assembly on the
source body.
Attach the filaments with the two screws.
General Maintenance
(Figure
23).
6
99