Figure 8-10. Fault 8, Low Heater Temp
LOI
When Fault 8 is detected, the LOI displays the "O
comments and procedures in paragraphs 1 through 4 above.
Troubleshooting
Rosemount Oxymitter 4000
Temp Low" message. Refer to the
2
mV, then an Open
Heater fault will be displayed
3. Power down the electronics.
electronic assembly per "Ele
Assembly Replacement" in S
Maintenance and Service. U
measure the resistance acro
heater connector, J8.
4. If the heater is good, the read
approximately 70 ohms. If th
see "Heater Strut Replacem
Section 9: Maintenance and
LOI
When Fault 8 is detected, the LO
Temp Low" message. Refer to the
procedures in paragraphs 1 throu
115