UNI-T UT219P User Manual page 16

Ac digital power clamp meter
Table of Contents

Advertisement

Note:
1) If the displayed power factor is not 1, please calculate the specification of power according
to the phase angle error.
2) [1P] 0.09kW~1000kW
3) [3P3W] 0.15kW~1732kW
4) [3P4W] 0.27kW~3000kW
3.2 Apparent power
Accuracy
Range
Resolution
15Hz~40Hz
40Hz~70Hz
Power factor: 1
Power factor: 1
±(2.0%+5)
60.00kVA
0.01kVA
600.0kVA
0.1kVA
±(2.5%+5)
±(1.7%+5)
1000kVA
1kVA
Note:
1) If the displayed power factor is not 1, please calculate the specification of power according
to the phase angle error.
2) [1P] 0.09kVA~1000kVA
3) [3P3W] 0.15kVA~1732kVA
4) [3P4W] 0.27kVA~3000kVA
3.3 Reactive power
Accuracy
Range
Resolution
15Hz~40Hz
40Hz~70Hz
Power factor: 1
Power factor: 1
±(2.5%+5)
60.00kVAr
0.01kVAr
600.0kVAr
±(3.0%+5)
0.1kVAr
±(2.0%+5)
1000kVAr
1kVAr
Note:
1) If the displayed power factor is not 1, please calculate the specification of power according
to the phase angle error.
2) [1P] 0.09kVAr~1000kVAr
3) [3P3W] 0.15kVAr~1732kVAr
4) [3P4W] 0.27kVAr~3000kVAr
3.4 Power factor
Accuracy
Range
Resolution
15Hz~1000Hz
Calculate the specification
-1~ 1
0.001
of power factor according
to the phase angle error
29
UT219P User Manual
Overload
protection
70Hz~1000Hz
Power factor: 1
1000A AC
1000Vrms
±(2.5%+5)
Overload
protection
70Hz~1000Hz
Power factor: 1
1000A AC
1000Vrms
±(3.0%+5)
Overload
Remark
protection
Min. measured
1000A AC
voltage: 30V
Max. measured
1000Vrms
current: 10A
UT219P User Manual
3.5 Phase angle
Accuracy
Range
Resolution
15Hz
40Hz
~
~
40Hz
70Hz
-180° (advance)
0.1°
±5°
±3°
~179.9° (delay)
Note:
1. The null-position cross phase difference between current and voltage waveforms is positive
if current lags behind voltage; negative if current is ahead of voltage.
2. An error of 2° is added if the measured conductor is not placed at the center of the clamp jaws.
4. Harmonic analysis
Function
Harmonic order
1
2~6
RMS of harmonic
7~8
of each order
9~10
11~15
16~30
30~40
Note:
1) Minimum measured voltage is 30V, minimum measured current: >10% of current range.
2) THD-F and THD-R are displayed on the LCD.
3) Harmonic analysis is recorded in the clamp, and can be displayed and viewed on mobile
APP via Bluetooth module.
4) If fundamental frequency is 100Hz, the harmonic order reaches 40.
If fundamental frequency is >100Hz, the harmonic order reaches 15.
5. Frequency
Range
Resolution
15Hz~1000Hz
0.1Hz
Note:
Minimum measured voltage is 30V, minimum measured current: >5% of current range.
6. Phase detection
Range
Frequency
Function
Phase Detect
80V~1000V
40Hz~80Hz
Note:
The LCD shows "1 2 3" to indicate positive phase sequence or "3 2 1" to indicate negative one.
The LCD shows "— — — —" to indicate phase loss or unmeasurable.
Overload
Remark
protection
70Hz
Min.
~
1000Hz
measured
1000A AC
voltage: 30V
Max.
1000Vrms
±5°
measured
current: 10A
Accuracy
Overload protection
15Hz~400Hz
±(3.0%+10)
±(3.5%+10)
1000A AC
±(4.5%+10)
1000Vrms
±(5.0%+10)
±(7.0%+10)
±(10.0%+10)
±(20.0%+10)
Overload
Accuracy
protection
15Hz~1000Hz
1000A AC
1000Vrms
±(0.3%+3)
Overload protection
1000Vrms
30