9 . Maintenance Chart:

Service and Maintenance of the Invacare
personnel who have received the necessary instruction or training:
Bed Serial no . __________
Date:
6 Monthly
12 Monthly
Electrical System:
a. The electrical system in this bed is a sealed unit and must be replaced.
b. A battery change can be performed by certified maintenance staff and requires the
removal of four screws securing the box to the bed frame and the withdrawal of a plug
from the control box.
Toe-Operated Castors .
Toe operated castors are the non-adjustable type, however they can quickly and easily be
replaced by certified maintenance staff.
Central Lock Castors .
Central Lock Castors are adjustable. Certified maintenance staff can tighten brakes simply by
tightening the grub screw protruding from the brake pad at the rear of the castor just above
the wheel.
Upper Posture (Fowler) Removal and Replacement .
Removal
1. Disconnect bed from mains supply.
2. Remove bolt-fastening fowler Rocker Arm (25mm solid square MS, mild bend one
end, 4 lugs attached) to Fowler Frame. (Leave rocker arm attached to actuator)
All information is subject to change without prior notice. Photographs are for illustration purposes only and may vary from product description.
Invacare® is a registered trademark and Yes, you can.™ is a trademark of Invacare® Corporation. 2010 Invacare Asia Pacific
1600 User's Manual
Correct operation of bed
Castor fittings tightened
Castor brakes checked
Actuators checked for visible damage
Actuators shafts clean
Wiring correct and undamaged. Plugs
undamaged and no interference with bed
movement
Damaged paintwork repaired
Check all pivot points
13
1600 bed must be performed by
®
Check
Initial