Using Pointer Graph Test Function - Agilent Technologies OmniBER 718 SDH User Manual

Table of Contents

Advertisement

Selecting Test Features

Using Pointer Graph Test Function

Using Pointer Graph Test Function

Pointer Graph shows the relative offset during the measurement period. This allows
the time relationship of AU or TU pointer movements to be observed. Up to 4 days
of storage allows long term effects such as Wander to be observed. If an alarm
occurs during the measurement period, a new graph starts at the centre of the display
(offset zero) after recovery from the alarm.
The Pointer Graph display can be logged to the chosen logging device. See
"Logging on Demand " page 261.
TIP:
The graph can also be viewed on the
end of the measurement.
1 Set up the receive SDH interface and payload as required. See "Setting SDH
HOW TO:
Receive Interface" page 49.
2 On the RECEIVE Test Function page, select
CAPTURE INTERVAL required.
The capture interval determines the time between captures. Low values of capture
interval should be chosen when a high degree of pointer movements is expected.
High values of capture interval should be chosen when a low degree of pointer
movements is expected, for example Wander over 1 day, use 5 MINS and Wander
over 4 days, use 20 MINS.
If, during a long term measurement (4 days), an event occurs at a particular time
each day, then the instrument can be set to log the results graph of that event.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SDH RESULTS
RESULTS
then choose the
PTR GRAPH
display at the
95

Advertisement

Table of Contents
loading