Selecting and Using "Other" Features
Trigger Output
Trigger Output
Description
Use the pulse output from the TRIG OUT port on the OmniBER Multirate
Analyzer module to trigger external equipment. A pulse is output whenever the
condition selected in the TRIGGER SOURCE field occurs.
How To
1 Select
2 Set the TRIGGER SOURCE filed as required from the following:
NONE
TX FRAME
RX FRAME
RX B1 (B1 error)
RX B2 (B2 error) or
RX B3 (B3 error)
Pulse Format
TTL Levels. Termination can be 75 ohm or 10 kohm.
RZ with active high pulse of 60usec (Nominal). Note that a single pulse is generated
if a frame contains one or more errors giving a maximum frequency of 8kHz.
Connector
BNC.
324
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OTHER
, and set FUNCTION to
.
TRIGGER OUTPUT