Siemens NIC2-2 Installation And Startup Manual page 42

Network id camera
Hide thumbs Also See for NIC2-2:
Table of Contents

Advertisement

NIC2-2 · Installation and Set-up Manual
5. Error Messages
Although all error messages are plain text and self explaining, some guidelines on how to
react and handle on these will be appropriate. The error message will disappears when any
key is pressed.
The following is a list of all error messages while in set-up mode:
5.1 Illegal value
An erroneous value has been entered. The value of the length and position of a field has an
upper limit.
Action:
Enter correct value.
5.2 The adjustment may be max 32
The vertical adjustment has a limit of 32 line-pixels.
Action:
If there is a need for a greater adjustment than 32 pixel-lines or a negative
adjustment (move the image upwards) the whole picture must be moved
manually, field by field.
5.3 Out of storage for FTEXT
Maximum numbers of F-texts are 30. These can be distributed between the two windows,
C1 and C1N, after one's own preference.
Action:
Use F-texts more efficiently. If there is two F-texts on the same line it is pos-
sible to merge them. The F-text limit is number of F-text not the total number
of F-text characters.
5.4 Length must be between 1 and 8
A Menu-text can be from one up to eight characters long.
Action:
Enter correct value. To exit without entering any value, press ESC-key.
5.5 Communication error
An interference of a setup up-/download has occurred.
Action: Select a lower speed and restart the communication session.
5.6 Unknown programming error
Action:
Restart the camera by switching it off and then on again.
Note! Changes made in set-up may not be saved.
5.7 Wrong Dicom Licence Code!
The Dicom protocol is activated and none or an invalid licence key number has been en-
tered.
Action:
Enter correct licence key or set "Dicom" to Disabled.
Triacon AB Sweden
33
Publication no. 201101
October 2005

Advertisement

Table of Contents
loading

Table of Contents