Settings Win1/Win2 - Siemens NIC2-2 Installation And Startup Manual

Network id camera
Hide thumbs Also See for NIC2-2:
Table of Contents

Advertisement

Publication no. 201101
3.2.3

Settings Win1/Win2

If this entry is selected, the following will be displayed
Settings
Status
Exposure:
Name:
The cursor is located at "Status" and can be moved between the different settings by press-
ing up- and down-arrow keys. If the softkeys Previous and Next are visible it is two or more
predefined values that can be selected by pressing the corresponding function keys F2 and
F3.
The following settings can be performed:
Toggle between Used/Not used (i.e. enable/disable the picture). If one pic-
Status
ture is switched to "Not used" the other one will automatically be switched to
"Used". A "Not used" picture can not be accessed by the operator. Factory
default is Used.
Basic exposure time in milliseconds. A value between 10 and 65535 can be
Exposure
set. Factory default is 100ms.
The actual exposure time is the basic exposure time set in this window mul-
tiplied by a factor determined by the exposure correction set in the exposure
window that is opened from the patient data window. The correction factor is:
The name of this picture. Can be edited to any name with a maximum length
Name
of 15 characters. Factory default is "Win1" and "Win2".
Toggle between Normal/Inverted. Determine whether the picture should
Text
show white text on black background or black text on white background. Fac-
tory default is Normal.
Toggle between Bottom/Top. Determine whether the picture should be print-
Location
ed on bottom or top of the film. Factory default is Bottom.
NOTE! If a window is set-up for a MIN-R2 cassette bottom marking and then
is changed to top marking the lower 16 pixel-lines of the image will be placed
outside the cassette window. This means that the whole image must be
moved 16 pixel-lines up, field by field, to fit the MIN-R2 cassette window. If a
Standard C1 cassette is used, no adjustments is needed when swapping
from bottom to top marking.
Offset between operator- and exposure display. The picture which will be
Vert adj
printed on the film can be adjusted down. Maximum adjustment is 32 pixel-
lines. Factory default is 0.
Setting if AP or PA projection should be default. Even if there is no AP/PA
Def proj
field defined, the picture will be mirrored according to this setting.
October 2005
:
Used
100ms
Previous
1
2
3
0.5
0.6
0.71
NIC2-2 · Installation and Set-up Manual
Text
: Normal
Location: Bottom
Vert adj:
0
Def proj: AP
Next
Save
4
5
6
0.84
1
1.19
8
7
8
9
1.44
1.68
2