Introduction; Setup For Diagnostic Loopback Tests; Diagnostics Methodology - Multitech MultiDSU MTT1DSU User Manual

Data service unit
Hide thumbs Also See for MultiDSU MTT1DSU:
Table of Contents

Advertisement

Introduction

MultiDSU diagnostics involve using the Diagnostics menus that are
accessible through option 4 on the Main Menu. Option 4 opens the
Diagnostics Menu that enables you to run loopback diagnostic tests on
the MultiDSU and its connecting lines.

Setup for Diagnostic Loopback Tests

Connect a T1 crossover cable between the LINE jacks on two
MultiDSUs. On the T1 Line Configuration menu, set up both units for the
identical framing format (option 1) and line coding (option 2). Use the
Clocking menu to set up one unit for Internal (option 1) (or External,
option 3, if your DTE is providing the 1.544 MHz clock), then set the
other unit for Line (option 2). Run the appropriate loopback tests from
the Diagnostics menu.

Diagnostics Methodology

The methodology for all the loopback tests is as follows:
1 Connect the DTE device to the Data port of the local MultiDSU.
2 Put the MultiDSU (either the local or remote) into the desired test
mode.
3 Send data from the DTE device (local or remote).
4 Verify that the data received from the MultiDSU matches exactly
what was sent.
5 If the data received matches the data that was sent, then the test
passed. If the data received is corrupted or no data is received,
then the test failed. A failed test indicates a problem with either the
cables, connections, or one of the units involved in the test. If the
cause of the failure is not readily apparent, try using the process of
elimination to isolate the trouble spot(s). If there is still no clear
cause of the problem, follow the steps provided in Chapter 6 to
contact Tech support.
42

Advertisement

Table of Contents
loading

Table of Contents