Diagnostics and troubleshooting
9.4 Status messages according to NE 107
Code
Text message
no measured value
available
F017
Adjustment span too
small
F025
Error in the
linearization table
F036
No operable software
F040
Error in the
electronics
F080
General software
error
F105
Determine measured
value
F260
Error in the calibration
F261
Error in the
transmitter settings
F265
Measurement
function disturbed
Function check
Code
Text message
C700
Simulation active
34
Cause
Transmitter tilted
Adjustment not within specification
Index markers are not continuously
rising, for example illogical value pairs
Checksum error if software update
failed or aborted
Limit value exceeded in signal
processing
Hardware error
General software error
The transmitter is still in the switch-on
phase, the measured value could not
yet be determined
Checksum error in the calibration
values
Error in the EEPROM
Error during setup
Auto false echo suppression faulty
Error when carrying out a reset
Program sequence of the measuring
function disturbed
Cause
A simulation is active
Rectification
Clean the antenna system
Change adjustment according to the
limit values (difference between min.
and max. ≥ 10 mm)
Check linearization table
Delete table/Create new
Repeat software update
Send transmitter for repair
Restart transmitter
Send transmitter for repair
Restart transmitter
Wait for the end of the switch-on
phase
Duration up to 3 minutes depending
on the measurement environment
and parameter settings
Send transmitter for repair
Repeat setup
Carry out a reset
Device restarts automatically
Rectification
Finish simulation
Wait for the automatic end after
60 mins.
Operating Instructions, 10/2021, 62705-76-01