Safety Standards; Hazardous Locations - National Instruments USB-9237 User Manual And Specifications

4-channel, 24-bit half/full-bridge analog input device
Table of Contents

Advertisement

For UL and other safety certifications, refer to the product label, or visit
Note
certification
in the Certification column.
Environmental
NI USB-9237 User Guide and Specifications
Isolation
Channel-to-channel..........................No isolation between channels
Channel-to-earth ground
Continuous ...............................60 VDC,
Withstand .................................1,000 V
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics. Do not
connect the NI USB-9237 to signals or use for measurements within
Measurement Categories II, III, or IV.

Safety Standards

The NI USB-9237 is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 612010-1
, search by model number or product line, and click the appropriate link

Hazardous Locations

The NI USB-9237 is not certified for use in hazardous locations.
The NI USB-9237 device is intended for indoor use only.
Operating temperature
(IEC60068-2-1, IEC 60068-2-2) ............0 to 60 °C
Storage temperature
(IEC60068-2-1, IEC 60068-2-2) ............–40 to 85 °C
Ingress protection ...................................IP 40
Measurement Category I
dielectric withstand test
20
, verified by a 5 s
rms
ni.com/
ni.com

Advertisement

Table of Contents
loading

Table of Contents