ResMed AirMini User Manual page 26

Hide thumbs Also See for AirMini:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
• The use of accessories other than those specified for the
machine is not recommended. This may result in increased
electromagnetic emissions or decreased immunity of the
machine and result in improper operation.
• The machine should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the
machine should be observed to verify normal operation in the
configuration in which it will be used.
• The AirMini mask systems and connectors are fitted with vents.
The full face connector for AirMini is also fitted with an AAV.
The AAV and vents have specific safety functions to prevent
carbon dioxide build-up in the mask. The AAV and vents should
be kept clean and not covered or blocked. The connectors
should not be used if the vent or AAV are damaged as they will
not be able to perform their safety functions. The connectors
should be replaced if the valves of the vent or AAV are
damaged, distorted or torn.
• The machine is not intended to be operated by persons
(including children) with reduced physical, sensory or mental
capabilities without adequate supervision by a person
responsible for the patient's safety.
• The device has not been tested or certified for use in the vicinity
of X-ray, CT or MRI equipment. Do not bring the device within
13 ft (4 m) of X-ray or CT equipment. Never bring the device into
an MR environment.
24

Advertisement

Table of Contents
loading

Table of Contents