7.5
Canada
7.5.1
I6 Canada Intrinsically Safe
Certificate 1091070
Standards CAN/CSA C22.2 No. 0-10, CSA Std. C22.2 No. 25-1966,
Markings
7.5.2
K6 Canada Intrinsically Safe, Explosionproof, and Division 2
Certificate 1091070
Standards CAN/CSA C22.2 No. 0-10, CSA Std. C22.2 No. 25-1966, CSA
Markings
7.6
Europe
7.6.1
E1 ATEX Flameproof
Certificate FM12ATEX0065X
Standards EN 60079-0: 2012+A11:2013, EN 60079-1: 2014, EN
Markings
Specific Conditions of Safe Use (X):
1. See certificate for ambient temperature range.
2. The non-metallic label may store an electrostatic charge and become
a source of ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than
four joules.
Quick Start Guide
CAN/CSA C22.2 No. 94-M91, CAN/CSA C22.2 No. 157-92, CSA
C22.2 No. 213-M1987, CAN/CSA C22.2 No. 60079-11:14,
C22.2 No 60529-05
IS CL I, DIV 1 GP A, B, C, D when installed per Rosemount
drawing 00248-1056; CL I DIV 2 GP A, B, C, D; Type 4X,
IP66/68
Std. C22.2 No. 30-M1986, CAN/CSA C22.2 No. 94-M91, CSA
Std. C22.2 No.142-M1987, CAN/CSA C22.2 No. 157-92, CSA
C22.2 No. 213-M1987, CAN/CSA C22.2 No. 60079-11:14,
C22.2 No 60529-05
XP CL I/II/III, DIV 1, GP B, C, D, E, F, G when installed per
Rosemount drawing 00644-1059; IS CL I, DIV 1 GP A, B, C, D
when installed per Rosemount drawing 00248-1056; CL I DIV
2 GP A, B, C, D; Type 4X, IP66/68; Conduit Seal not required
60529:1991 +A1:2000 + A2:2013
II 2 G Ex db IIC T6...T1 Gb, T6(-50 °C ≤ T
T1(-50 °C ≤ T
≤ +60 °C)
a
See
Table 7-2
for process temperatures.