Fluke 189II/AN Service Manual page 30

True-rms digital multimeter
Table of Contents

Advertisement

189II/AN
Service Manual
Step
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
20
Table 4. 189II/AN Performance Test Steps (cont.)
Function
Range
DC mV
50.000 mV
Discontinue REL Offsets
DC mV
500.00 mV
DC mV
500.00 mV
mV DC/AC
500.00 mV
mV AC/DC
500.00 mV
Set 5520A for 2-Wire COMP and Connect in 4-Wire configuration
500.00 Ω
Ohms
500.00 Ω
Ohms
500.00 Ω
Ohms
Discontinue REL Offsets
Ohms
5.0000 kΩ
Ohms
50.000 kΩ
Set 5520A for COMP OFF and Remove Auxiliary Test Cable
Ohms
500.00 kΩ
Ohms
5.0000 MΩ
Ohms
30.000 MΩ
Ohms
500.0 MΩ
AC mV
50.000 mV
AC mV
50.000 mV
AC mV
500.00 mV
AC mV
500.00 mV
AC mV
500.00 mV
AC mV, Hz
500.00 mV
AC mV, Hz
500.00 mV
VAC
5.0000 V
VAC
5.0000 V
VAC
5.0000 V
VAC, HZ, %
5.0000 V
(Duty Cycle)
VAC
50.000 V
Frequency
Input Level
or Model
[2]
50 mV
500 mV
-250 mV
50 mV
0 Hz
250 mV
35 kHz
0 Ω
REL Offset
[2]
0.2 Ω
[2]
500 Ω
5 kΩ
50 kΩ
500 kΩ
5 MΩ
30 MΩ
300 MΩ
5 mV
20 Hz
50 mV
65 kHz
50 mV
100 kHz
250 mV
65 kHz
500 mV
45 Hz
500 mV
45 Hz
600 mV
950 kHz
0.1 V
60 Hz
0.5 V
10 kHz
3 V
100 kHz
5 V p-p, Sq.
50 kHz
wave @15 %
15 V
100 kHz
[1]
Display Reading
Lower
Upper
Limit
Limit
49.955
50.045
499.85
500.15
-250.08
-249.92
49.97
50.03
237.10
262.90
-0.10
0.10
0.10
0.30
499.65
500.35
4.9973
5.0027
49.973
50.027
499.73
500.27
4.9921
5.0079
29.546
30.454
275.8
324.2
4.865
5.135
48.210
51.790
47.85
52.15
240.85
259.15
498.25
501.75
44.986
45.014
949.90
950.10
0.0952
0.1048
0.4945
0.5055
2.8160
3.1840
1.40
28.60
14.435
15.565