Liquid Sterilization (Usa Only); Autoclave Sterilization; Storage - Hitachi UST-676P Instruction Manual

Transrectum puncture electronic convex probe
Table of Contents

Advertisement

5-11-3. Liquid sterilization (USA only)

• Applicable chemical solution for sterilization
The applicable sterilants are listed below.
General name
Hydrogen peroxide
Peracetic acid
Glutaraldehyde
Glutaraldehyde
Glutaral
Note: * indicates that the marked sterilant is not applicable in Canada.
Warning
After chemical sterilization, thoroughly rinse the probe with sterile water.
Residual sterilant can cause an adverse reaction to the operator or patient.

5-11-4. Autoclave sterilization

A). Probe
Warning
The probe cannot withstand autoclave sterilization.
B). Accessories
Warning
The washing brush cannot withstand autoclave sterilization.

5-12. Storage

For details about the storage environmental conditions, see section 2-5-2 "Storage environmental conditions".
Trade name
*
PERASAFE
®
Practical liquid 1.62W/V%
Acecide
*
®
Solution 6%
WAVICIDE
-01 *
®
Solution 2.65%
STERANIOS *
Solution 2.0%
Cidex plus
®
Solution 3.4%
-47-
Manufacturer
ANTEC INTERNATIONAL
Saraya Co., Ltd.
Medical Chemical Corporation
Laboratoires ANIOS
ADVANCED STERILIZATION PRODUCTS
A Johnson & Johnson company
Division of Ethicon, Inc.
MN1-5029 Rev.22
®