3. Preparations for Use

3-1. Start up check of the probe

3-1-1. Visual check

Visually check the ultrasonic irradiation area, insertion portion, handle, cable, connector and rubber
boots.
If any holes, indentations, abrasion, cracks, deformation, looseness, discoloration, or other abnormalities
are found, do not use the equipment.

3-1-2. Verification of cleaning, disinfection and sterilization

Verify that cleaning, disinfection and sterilization are conducted according to the intended use.

3-1-3. Verification of operation

Connect to the ultrasound diagnostic instrument by following the instructions in section 4-6, "Connecting
to the ultrasound diagnostic instrument" and check that the selected probe match the convex display
and the displayed frequency and check the image for errors.
Remarks
For details on the displayed screens, see the documentation supplied with the ultrasound diagnostic
instrument.
If the probe is operated in still air, brightness on the top of the image may be non uniform, but this does
not affect the performance of the probe.
Warning
Be sure to preparations for use.
Using the probe and the puncture guide tube without noticing an abnormal condition can result in
injury to the operator or patient. If an inspection finds an abnormal condition in the probe, immediately
stop use and contact one of our offices and/or distributor's offices listed on the back cover.
Caution
Do not use the probe if the selected probe and image do not match the frequency.
An incorrect acoustic output can result in burns or other injuries to the patient. Contact one of our
offices and/or distributor's offices listed on the back cover.