Hardware States
System Parameter
System Phaselock
Mult. Source
Leveling Source #1
Leveling Source #2
Sweep Mode
Set source #2 power leveling to Ext. if an HP 8349B amplier is
1
used to amplify the LO signal.
Instrument States
Dene:
Source 1 Mult. Numer. = 1 Mult. Denom. = 1 Oset Frequency = 0.000000000 GHz
Source 2 Mult. Numer. = 1 Mult. Denom. = 1 Oset Frequency = 0.020000017 GHz
Receiver Constant Frequency = 0.020000000 GHz
All manuals and user guides at all-guides.com
During pulsed-RF
operation
None
On/Save
Int.
Int.
STEP or FRQ LIST
System Parameter
Pulsed-RF
Power Source #1
+8 dBm
Power Slope src 1
Power Source #2
10 dBm
Power Slope src 2
2.5 dB/GHz
Pulse cong detector wide BW
Set source #1 power level to the highest level possible
1
without IF overload. Power must be set in the 'normal
BW' mode then changed to the 'wide BW' mode, if
desired.
Set source #2 power to -10 dBm if an HP 8349B
2
amplier is used to amplify the LO signal
Set to normal BW for nonpulsed-RF operation
3
Settings:
During operation
of 2nd test set
Int.
O
Int.
N/A
1
Any
CW
10 dBm
1
OFF
OFF
N/A
2
N/A
normal BW
3