HP 16532A Service Manual page 70

1 gsa/s digitizing oscilloscope
Table of Contents

Advertisement

HP 16532A - Service
5. On the HP 54502A Oscilloscope, set the following parameters, in the order given.
Menu
Channel
Timebase
Trigger
6. In the HP 16532A Oscilloscope trigger menu, select Auto-Trig On. Verify that the 100 MHz System
Clock is gated.
7. Select STOP on both the HP 54502A and HP 16532A oscilloscopes.
8. Disconnect the cable from both oscilloscopes.
6-12
Selection
Probe
V/Div
Offset
Coupling
s/Div
Acquisition
Mode
Level
Slope
Setting
1.000:1
200 mV
0.0 V
DC
10 ns
Realtime
Edge
− 200 mV
Falling

Advertisement