HP 16532A Service Manual page 13

1 gsa/s digitizing oscilloscope
Table of Contents

Advertisement

TRIGGERING
Trigger Level Range:
Trigger Modes:
Immediate:
Edge:
Pattern:
Auto-Trigger:
Events Delay:
Intermodule:
External:
DIGITIZER
Resolution: 8 bits (1 part in 256)
Digitizing Rate: Up to 1 Gigasample per second
Digitizing Technique: Real-time digitizing; each 8000 samples are acquired on a single acquisition
Acquisition Memory Size: 8000 samples per channel
Table 1-2. HP 16532A Operating Characteristics (cont.)
Within display window (full scale and offset)
Triggers immediately after arming condition is met.
Triggers on rising or falling edge of any internal channel or external trigger,
count adjustable from 1 to 32,000.
Triggers on entering, exiting, duration greater than, duration less than, and
duration in range for a specified pattern of both internal channels and external
trigger, count adjustable from 1 to 32,000. Duration time range is adjustable
from 20 ns to 160 ms in 10 ns steps. Duration accuracy is approximately ± 3%
± 2 ns.
If enabled, the module will self-trigger if no trigger condition is found within
approximately 50 ms after arming.
The trigger can be set to occur on the nth edge or pattern, as specified by the
user. The number of events (n) can be set from 1 to 32,000 events. Maximum
count frequency is 70 MHz.
Arms another measurement module or triggers the rear panel BNC.
If enabled, the oscilloscope will trigger on an ECL level signal (-1.3 V).