The following checklist provides an overview of the procedures necessary to replace the hardware 
contained in this kit.
❑ Remove the electric actuator.
❑ Remove the liftarm assembly.
❑ Install liftarm assembly.
❑ Install the electric actuator.
Removing/Installing the Electric Actuator
NOTE: For this procedure, refer to FIGURE 1 on page 3.
1. Unplug the electric actuator from the control box on the mast assembly. Thread the actuator lead through 
the actuator lead routing hole in the mast assembly.
2. Remove the hardware that secures the top of the electric actuator to the lift arm actuator mounting 
bracket.
A. Remove the rubber caps from the nut and screw that attach the electric actuator to the lift arm 
actuator mounting bracket.
B. Remove the screw, washer, bearing, nylon washer, actuator cap, bushing, nylon washer, bearing, 
washer, flat washer and nut.
NOTE: The actuator cap will become loose when the attaching hardware is removed. Inspect the actuator cap for 
damage. Replace the actuator cap if damage is found.
3. Remove the hardware that secures the bottom of the electric actuator to the mast assembly actuator 
mounting bracket.
A. Remove the rubber caps from the nut and screw that attach the electric actuator to the mast assembly 
actuator mounting bracket.
B. Remove the screw, washer, bearing, nylon washer, bushing, nylon washer, bearing, washer, flat 
washer, and nut.
4. To install the electric actuator, reverse STEPS 1‐3.
Roze™ Patient Lift Liftarm Assembly
Hardware Replacement Kit
2
Part No 1160830