P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
8-2-4
Rear Cover Replacement Procedure
8-2-4-1
Tools
Phillips screwdriver of appropriate size.
8-2-4-2
Time Required
5 min
8-2-4-3
Preparations
1.) Shut down the Venue Fit™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
8-2-4-4
Rear Cover Removal Procedure
1) Remove the battery.
•
2) Release 7 captive screws.
Note:
IMPORTANT! Ensure the screws are fully released before removing the cover, to avoid
damaging the touch display screw heli-coils.
Battery Removal Procedure
Figure 8-36 Releasing 7 captive Screws
Chapter 8 - Replacement Procedures
VENUE FIT™ SERVICE MANUAL
Power
Shutdown.
Mounting / Dismounting System on Cart