Rear Cover & Latch Assembly Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-2-5
Rear Cover & Latch Assembly Replacement Procedure
NOTE:
This section describes the replacement procedure for Rear Cover & Latch Assembly - supplied as part
of the Cover Kit (Vivid i n systems Part No. S2423361; Vivid q N systems Part No. S2423362).
8-2-5-1
Tools
None required.
8-2-5-2
Time Required
30 minutes
8-2-5-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-5-4
Rear Cover & Latch Assembly Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Remove the Bearing Handle as described in the
3) Remove the Control Panel and Keyboard as described in the
Procedure on page 8 -
4) Remove the LCD Display Frame as described in the
page 8 -
5) Remove the LCD Display; see the
6.) Loosen and remove the two screws on each of the left and right hinges (attached to the Rear Cover
and Latch Assembly), as shown in
7.) Lift the Rear Cover and Latch Assembly away from the Bottom Assembly and remove completely
from the system.
8-2-5-5
Rear Cover & Latch Assembly Installation Procedure
1.) Return the Rear Cover and Latch Assembly to the system and place it in position on the Bottom
Assembly, carefully checking that the left and right hinges are properly aligned.
2.) Return and fasten the two screws on each of the left and right hinges (previously removed - refer
to
Figure
3) Return the LCD Display; see the
4) Return the LCD Display Frame as described in the
page 8 -
1
EVISION
9.
27.
Figure 8-23 Removing the Hinge Screws
8-23).
28.
Chapter 8 - Replacement Procedures
Bearing Handle Removal Procedure on page 8 -
LCD Display Frame Removal Procedure on
LCD Display Removal Procedure on page 8 -
Figure 8-23
below.
LCD Display Installation Procedure on page 8 -
LCD Display Frame Installation Procedure on
V
V
N S
IVID I N AND
IVID Q
Control Panel and Keyboard Removal
53.
55.
M
ERVICE
ANUAL
3.
8-19

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid q nFl000110Fl000090Fq000001Vivid-iVivid-q