Air Filter Cleaning

Lock the Ultrasound system's wheels prior to removing/cleaning the air filter. This prevents the
CAUTION
system from moving unexpectedly.
Table 10-9
Step
1
2
3
NOTE:
For convenience (or if the air filter is excessively dirty), replacement filters are available. Refer
to
Table 9-8
1
EVISION
Air Filter Cleaning - Frequency Varies with Your Environment
Item
Remove the left side cover from the system, as described in
Remove Filter
Removal Procedure on page 8 -
Cover
Remove the air filter from the cover, as shown in
The filter can be cleaned in sprinkling water, or it can be dusted with a vacuum cleaner. If
the filter is metal, wash and/or vacuum. If the filter is fiber or plastic, vacuum or replace. Dry
Clean Filter
the filter.
Install the clean (and dry) filter. Refer to the
Install Filter
air filter installation instructions.
on page 9-7 for the air filter replacement part number.
Chapter 10 - Care and Maintenance
V
S60N/V
S70N
IVID
IVID
Description
Air Inlet (Left Side) Cover
7.
Figure 8-4 on page
Air Filter Installation Procedure on page 8 - 9
B
S
M
ASIC
ERVICE
ANUAL
8-8.