Maintenance Method; Daily Inspection And Maintenance - Yamato BS 200 Instruction Manual

Water bath
Table of Contents

Advertisement

Daily Inspection and Maintenance

WARNING!
Disconnect the power cable from the power source when doing an inspection or maintenance
unless needed.
Perform the daily inspection and maintenance after returning the temperature of this unit to the
normal one.
Do not disassemble this unit.
CAUTION!
Use a well-drained soft cloth to wipe dirt on this
unit. Do not use benzene, thinner or cleanser for
wiping.
Do not scrub this unit.
deterioration or color change may result in.
Monthly maintenance
Check the earth leakage breaker function.
1. Connect the power cord.
2. Turn the breaker on.
3. Push the red test switch by a ballpoint pen etc.
4. If there is no problem, the earth leakage breaker will be turned off.
For any questions, contact the dealer who you purchased this unit from, or the nearest sales
division in our company.
Deformation,